hplcAnalyzer turns a peptide HPLC-UV chromatogram into a concentration and a purity number without a standard curve. It reads the raw instrument files, corrects the baseline, integrates the peaks, and converts the main peak area to molarity using a molar absorptivity predicted from the peptide sequence. Everything runs locally, either from the R console or from a Shiny app aimed at bench chemists rather than R programmers.
Version 0.7.8. See NEWS.md for the version history.
Three things people actually use this for. Each is a complete recipe; the walkthrough below shows the clicks.
You have sixty .D folders off the prep queue and you need a percent-purity number for each.
Point the app at the folder, click Load samples, read the Area (%) column of the peak
table, download the CSV. Purity is an area ratio, so it needs no extinction coefficient and no
sequence: it works even for peptides the app cannot quantify. Set Max analyte RT to 80
percent first, or the column regeneration step at the end of the run counts as a peak and
eats into the denominator.
You need micromolar for each peptide, to normalise a set of stocks. Same batch, but now every
run needs a sequence, which the app resolves from the folder name, SAMPLE.XML, or a CSV map
you upload. Set the injection volume if the method is not 100 microlitres. Read Conc final
(µM). There is no calibrant anywhere in this: the absorptivity is predicted from the
sequence, so one injection per peptide is enough.
A concentration looks wrong and you want a second opinion that does not involve running
anything again. If the peptide contains a tryptophan or a tyrosine, switch the wavelength to
280 nm and re-load: the same injection is quantified through a completely different
chromophore and a different absorptivity model. On the batches behind
Recalibration of eps214 the two channels agree to a median of 8
percent, so a peptide where they disagree by 30 percent is telling you something. Peptides
with neither Trp nor Tyr return NA (missing ε) at 280 nm, by design, and cannot be
cross-checked this way.
| Instrument | Files read | Reader |
|---|---|---|
| Agilent ChemStation | a folder of .D acquisition folders |
read_hplc_agilent(), via chromConverter |
| Thermo / Chromeleon | *_UV_VIS_N.txt UV text exports |
read_hplc_thermo() |
Both readers are found automatically in whichever folder you point the app at. Agilent .D
folders are picked up as immediate subdirectories; Thermo exports are picked up as files
directly inside the folder. Neither search recurses.
A real wavelength choice exists for Agilent only. For an Agilent .D folder,
read_hplc_agilent() searches the DAD signals for one whose detector_range attribute
matches both the requested signal wavelength and the reference wavelength, so asking for
280 nm genuinely extracts the 280 nm trace. The Thermo reader has no wavelength concept at
all: read_hplc_thermo(file) takes a path and nothing else, and the app's "Thermo UV channel"
control selects a file number (N in *_UV_VIS_N.txt), not a wavelength. On the Thermo
path the wavelength radio button therefore only chooses which epsilon formula is applied to
whatever trace the channel number pulled in.
The package is pure R. No compiler, and no Rtools on Windows, is needed for hplcAnalyzer itself, though some of its CRAN dependencies compile from source on Linux.
DESCRIPTION declares no minimum R version, but chromConverter, dplyr and ggplot2 all
require R 4.1 or newer, so that is the effective floor. Developed and tested on R 4.5.2.
1. Install R. hplcAnalyzer needs R 4.1 or newer; 4.5 is what it is developed on.
| Platform | How |
|---|---|
| macOS | Download the .pkg from cran.r-project.org/bin/macosx. Take the arm64 build on Apple Silicon (M1 and later) and the x86_64 build on Intel. Double click, accept the defaults. |
| Windows | Download the installer from cran.r-project.org/bin/windows/base. Accept the defaults. Rtools is not needed. |
| Ubuntu / Debian | sudo apt update && sudo apt install r-base r-base-dev. The -dev package matters: several CRAN dependencies compile from source on Linux. |
RStudio is optional and changes nothing here. If you use it, install it after R.
2. Get the code. Either clone it:
git clone git@github.com:PitKubi/hplcAnalyzer.git
cd hplcAnalyzeror download the ZIP from the repository's green Code button and unzip it. On macOS the
unzipped folder lands in ~/Downloads/hplcAnalyzer-main; cd there.
3. Install the package and everything it needs, in one command:
Rscript install.RExpect five to fifteen minutes on a first run, longer on Linux where packages compile. It is safe to re-run; anything already present is skipped.
4. Start the app.
Rscript -e 'hplcAnalyzer::run_hplc_app()'Your browser opens on the app. Leave the terminal window open, it is the server. Ctrl-C in that terminal stops it.
If step 3 fails, the usual cause on macOS is a package that wants to compile. Force binary installs and try again:
options(pkgType = "binary")
source("install.R")On Ubuntu a compile failure usually names a missing system library. sudo apt install libcurl4-openssl-dev libssl-dev libxml2-dev libfontconfig1-dev covers the ones these
dependencies ask for.
git clone git@github.com:PitKubi/hplcAnalyzer.git
cd hplcAnalyzer
Rscript install.Rinstall.R reads the dependency list out of DESCRIPTION, installs from CRAN whatever is
missing, then installs the package from the working tree. It is written in base R so that a
machine with nothing but R on it can run it. It also works from inside R or RStudio:
setwd("/path/to/hplcAnalyzer")
source("install.R")If the repository is public, this is all it takes:
install.packages("remotes")
remotes::install_github("PitKubi/hplcAnalyzer")While it is private, that call fails with a 404 rather than a 403, because GitHub hides
private repositories rather than refusing them. Use a personal access token with the repo
scope:
install.packages("remotes")
remotes::install_github("PitKubi/hplcAnalyzer",
auth_token = Sys.getenv("GITHUB_PAT"))Put the token in ~/.Renviron as GITHUB_PAT=ghp_... rather than typing it into a script.
Never commit it.
The repository was renamed from hplc_analyzer to hplcAnalyzer on 1 September 2026, so that
the repository and the package are the same word. GitHub redirects the old address permanently,
so anything already pointing at PitKubi/hplc_analyzer keeps working.
This is how the package currently reaches collaborators on macOS and Windows. Build it once on any machine:
cd hplcAnalyzer
R CMD build .That writes hplcAnalyzer_0.7.8.tar.gz. Send that file. The recipient installs the CRAN
dependencies once and then the tarball:
install.packages(c("chromConverter","dplyr","baseline","signal","pracma","ggplot2",
"gridExtra","shiny","shinyFiles","fs","DT","tibble","xml2","magrittr"))
install.packages("C:/path/to/hplcAnalyzer_0.7.8.tar.gz", repos = NULL, type = "source")On Windows, close and reopen R before installing over an existing version. A loaded package cannot be overwritten and the install fails quietly. Check what you ended up with:
packageVersion("hplcAnalyzer")INSTALL.md carries the same instructions in a longer, step by step form for users who are
new to R.
library(hplcAnalyzer)
run_hplc_app()The app opens in your browser. Nothing leaves your machine: there is no upload, no account and no network call anywhere in the package.
The run shown below is a real Agilent batch of four peptide injections, chosen because each main peak resolves to baseline on both sides; see how a peak is integrated for what happens when it does not. The peptide names are synthetic, so the numbers belong to the sequences on screen rather than to anyone's samples. There is no blank in this folder, so the screenshots show the plain ALS baseline path; see the blank-subtracting path leaves an injector artefact for what changes when a blank is present.
Step 1. Point it at a folder. Click Choose folder directory and pick the folder that
holds your .D folders, or your *_UV_VIS_N.txt exports. Pick the folder itself, not one of
the .D folders inside it.
Step 2. Set the wavelength and the injection volume, then load. 214 nm is the default and is the one that works for every peptide. For Agilent, set the injection volume if the method is not 100 microlitres, because concentration scales inversely with it. Then click Load samples: every run is analysed once and the table fills in.
Each row carries the retention time and area of the main peak, the concentration before and after the dilution factor, the sequence that was resolved, and a status. A blank, if one is present in the folder, is detected and used rather than listed as a sample.
Step 3. Inspect one run. Click a row, or use Previous and Next. The top plot shows the raw trace, the fitted baseline and the corrected trace, so you can see what the baseline correction did rather than trust it. The middle plot shows the whole acquired run with the integrated peak shaded and the numbers that produced the concentration, including the epsilon actually used. The bottom plot, Integration detail, is the same peak zoomed: the orange chord is the baseline the area was measured against, the shaded region is what was counted, and a dotted line marks any point where a fused neighbour was cut away. The run above it is never truncated, which is why the detail gets its own panel. The sidebar fills with the top ten peaks and their Area (%).
Step 4. Cross-check at 280 nm, where the peptide allows it. Switch the wavelength and
click Load samples again. Peptides carrying a tryptophan or a tyrosine are quantified
through a second, independent chromophore. Peptides carrying neither return NA (missing ε)
and say so in plain words rather than reporting a number that would be meaningless.
In this batch, SAMPLEWK, SAMPLEWWK and SAMPLEPEPTIDEY quantify at both wavelengths;
TESTPEPTIDEK has no 280 nm chromophore at all and returns NA (missing eps). That is not a
failure, it is the reason 214 nm is the default: only about a third of tryptic peptides carry
Trp or Tyr. Note also how much smaller the 280 nm signal is, and how much it depends on what
the peptide carries: on the same four injections the main peak area at 280 nm is 43.74, 113.94,
7.23 and 0.38 mAU-min against 215.39, 455.38, 122.10 and 166.61 at 214 nm. Two tryptophans
give a usable 280 nm peak; a single tyrosine gives one seventeen times smaller.
Step 5. Download the results. Download Results CSV writes one row per run; the columns are listed under Results CSV columns.
library(hplcAnalyzer)
result <- run_hplc_analysis_agilent(
sample_d_path = "batch/004-P2-A1-MRMP-00000001-001_SAMPLEPEPTIDEK.D",
blank_d_path = "batch/003-1-BLANK0.D",
peptide_sequence = "SAMPLEPEPTIDEK",
use_hybrid = TRUE,
signal_wavelength = 214,
inj_vol_ml = 0.1,
min_rt_frac = 0.30,
max_rt_frac = 0.80
)
result$peak_table # one row per detected peak
result$epsilon # molar absorptivity used, M^-1 cm^-1
result$concentration_uM # main peak concentration, micromolar
result$plot # annotated ggplot
main_peak_area_percent(result$peak_table) # purity, percentrun_hplc_analysis_thermo() takes sample_file instead of sample_d_path and accepts no
blank.
Both pipelines run the same steps in the same order.
- Read.
read_hplc_agilent()orread_hplc_thermo()returns a data frame oftime(min) andintensity(mAU). The Agilent reader also parses the injection volume out ofacq.txtand stamps it on the frame as theinj_vol_mlattribute. - Baseline correction.
baseline_als()fits an asymmetric least squares baseline to the whole trace and subtracts it. This is the default and the only path at 280 nm.align_subtract_then_hybrid()is the blank-subtracting path, used at 214 nm when a blank run is available. It aligns the blank to the sample on the injector peak midpoint, subtracts it only after the injector window, bridges the positive/negative artefact pairs that over-subtraction creates, then callsbaseline_hybrid_sm()to fit piecewise ALS baselines between the minima of the difference trace. A one minute guard ramps the baseline up from zero just after the injector so the subtraction does not open with a dip.- Thermo exports are taken as already baseline corrected by the instrument, so
correctedis set equal tointensity.
- Peak detection.
detect_peaks_on_smoothed()smooths the corrected trace with a Savitzky-Golay filter (11 point window, 3rd order by default), estimates the noise as the standard deviation of the first 50 points, and callspracma::findpeaks()with a threshold ofsnrtimes that noise (default 5). Peaks outside the analyte retention time window are dropped. - Area. Trapezoidal integration (
pracma::trapz()) between the peak start and end indices thatfindpeaks()reports, on the smoothed trace. Units are mAU times min. - Concentration.
estimate_epsilon_214()orestimate_epsilon_280()predicts the molar absorptivity from the sequence, andcalculate_peak_conc()applies Beer-Lambert.
filter_top_peaks() ranks peaks by height and keeps the top ten; peak_area_percent() and
main_peak_area_percent() turn that ranked set into the Area (%) purity column.
Two functions cover 214 nm. published_epsilon_214() is the literature model, untouched.
estimate_epsilon_214() is what the pipelines and the app actually call, and it applies a
recalibration measured on our own data; see Recalibration of eps214
below for what changed, why, and how to change it back.
published_epsilon_214() implements the additive residue model of
Kuipers, B. J. H. and Gruppen, H. (2007). Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm to enable quantitative reverse phase high-performance liquid chromatography-mass spectrometry analysis. Journal of Agricultural and Food Chemistry 55(14), 5445-5451.
At 214 nm the peptide bond itself is the dominant chromophore, so absorptivity is built up term by term rather than from aromatic residues alone:
-
923 M-1 cm-1 per peptide bond, that is
nchar(sequence) - 1bonds. -
2675 M-1 cm-1 per proline that is not N-terminal. Proline is deliberately absent from the per-residue table below and carried by this term instead.
-
Per-residue contributions, summed over the sequence:
Residue eps Residue eps Residue eps Residue eps W 29050* H 5125 N 136 K 41 Y 5375 M 980 C 225 T 41 F 5200 R 102 G 21 E 78 Q 142 V 43 I 45 L 45 A 32 S 34 D 58 * Tryptophan is the one value
estimate_epsilon_214()does not use as printed. See Recalibration of eps214. -
Four tripeptides are special-cased to the measured values from the paper's Table 3 and bypass the additive model entirely: GGG 1080, GPG 3620, PGG 950, GGP 3880. They carry no tryptophan, so the recalibration scales them like any other tryptophan-free peptide.
Both functions return NA for anything that is not a string of the twenty standard one-letter
codes, so blanks, washes and unresolved sequences fall through cleanly.
The tryptophan term shipped here is not the published one. estimate_epsilon_214() returns
NON_TRYPTOPHAN_EPSILON_214_SCALE * (published - 29050 * nTrp) + MEASURED_TRYPTOPHAN_EPSILON_214 * nTrp
with MEASURED_TRYPTOPHAN_EPSILON_214 = 17340 and NON_TRYPTOPHAN_EPSILON_214_SCALE = 1.118.
Both constants sit at the top of R/estimate_epsilon_214.R, and the same two numbers appear in
peptide-calculator/src/calculations.js. Change them together or the desktop calculator and the
Shiny app will disagree.
Where the numbers come from. A peptide measured at both 214 and 280 nm gives two independent concentrations from one injection, so sample amount, dilution and any whole-batch offset cancel in their ratio and what is left is the extinction model. On 130 such injections that ratio tracks tryptophan and nothing else, at rho = -0.76, p < 1e-13: median 1.17 with no tryptophan, 0.82 with one, 0.68 with two. Tyrosine, phenylalanine, histidine, methionine and peptide length add nothing.
A reported concentration is inversely proportional to the extinction coefficient used, so the coefficient a peptide would have needed is its published value times that ratio. Fitting one replacement tryptophan term and one scale on everything else, by least squares over the 71 injections passing an 80 percent purity gate, with a 4000-draw bootstrap:
| Parameter | Fitted | 95 % CI | Published |
|---|---|---|---|
| eps214 tryptophan | 17,340 | 14,519 to 19,569 | 29,050 |
| scale on the rest | 1.118 | 1.01 to 1.26 | 1 |
Tyrosine and phenylalanine, freed in the same fit, come back at 5,080 and 3,813 with intervals covering their published 5,375 and 5,200, so they are left alone.
Why it is believable. Fitting each batch separately gives 15,300 and 18,200 for the tryptophan term, so it is not one batch misbehaving. Amino acid analysis, which is not used in the fit at all, agrees independently: tryptophan drags the 214 nm channel away from it while the 280 nm channel stays put. And the correction closes that gap too, from 26 to 11 percentage points in one batch and 20 to 7 in the other.
What it buys. Across the 130 injections, 214 nm against 280 nm improves from R2 0.68 to 0.87, and the median absolute difference between the two channels, as a percentage of the mean of the pair, from 17.3 to 8.1 percent. Those are the numbers the shipped constants actually produce, not the unrounded fit.
The scale is the soft number. The tryptophan term is well determined and its interval is nowhere near 29,050. The 1.118 scale is not: its interval nearly touches 1, and anchored on amino acid analysis instead, using the 170 tryptophan-free peptides of a clean batch, it comes out 1.078. Treat 1.08 to 1.12 as the range and set it from a batch you trust. Shipping 1 instead of 1.118 leaves the two channels 15 percent apart, so it is not a safe default; the two parameters belong together.
What changes in your numbers. Concentrations move as published / calibrated:
| Peptide | Trp | Published eps | Calibrated eps | Reported concentration |
|---|---|---|---|---|
DIAAYIK |
0 | 11,166 | 12,484 | 10.6 % lower |
GSEMVVAGK |
0 | 8,677 | 9,701 | 10.6 % lower |
INEWLTK |
1 | 34,974 | 23,963 | 45.9 % higher |
AWVNQLETQTGEASK |
1 | 42,878 | 32,800 | 30.7 % higher |
Every tryptophan-free peptide moves by the same 10.6 percent, so purity, Area (%) and any
ratio between peptides in one run are unaffected. Only tryptophan peptides change relative to
the rest, and they are the point of the exercise.
To go back to the literature model, either call published_epsilon_214() directly, or set
MEASURED_TRYPTOPHAN_EPSILON_214 <- 29050 and NON_TRYPTOPHAN_EPSILON_214_SCALE <- 1 and
reinstall. tests/testthat/test-estimate_epsilon_214.R pins the published model separately, so
it keeps passing either way.
The source data are a collaborator's purity batches and are deliberately not in this
repository. The 2025 batches in aaa_hplc_compare/ are the same comparison run earlier, kept
here in anonymised form; see the README in that folder.
estimate_epsilon_280() implements the Edelhoch method as revised by
Pace, C. N., Vajdos, F., Fee, L., Grimsley, G. and Gray, T. (1995). How to measure and predict the molar absorption coefficient of a protein. Protein Science 4(11), 2411-2423.
eps280 = 5500 * nTrp + 1490 * nTyr
Two things about this matter more than anything else in the package:
- Cystine is not included. The full Pace formula adds 125 M-1 cm-1 per disulfide bond. The disulfide state of a synthetic purity-check peptide is not known from its sequence, so this implementation assumes all cysteines are free and contributes nothing for them.
- A peptide with no Trp and no Tyr returns
NA, and cannot be quantified at 280 nm at all. There is no chromophore, so there is no concentration to compute, and the app saysNA (missing eps)rather than showing a number. In one production batch of 47 distinct peptides, 25 had neither Trp nor Tyr, so more than half the plate is unquantifiable at 280 nm. Use 214 nm unless you have a specific reason not to.
For those peptides the Area (%) purity column still works, because purity is a ratio of two areas and needs no absorptivity.
calculate_peak_conc() is Beer-Lambert with the units cancelled explicitly:
c [M] = (area [mAU*min] * flow [mL/min] / 1000) / (eps [M^-1 cm^-1] * pathlength [cm] * V_inj [mL])
The chain of units is:
area * flowgives mAU times mL, the total absorbance-volume swept through the cell.- Dividing by 1000 converts mAU to AU, giving AU times mL.
- Dividing by
eps * pathlengthconverts absorbance to molarity, leaving mol/L times mL. - Dividing by the injection volume in mL cancels the mL and leaves mol/L.
The result is multiplied by 1e6 for the reported micromolar value. Defaults are 1 mL/min flow and 1 cm pathlength; neither is exposed in the app, so change them in a script if your method differs.
Two datasets have been used to test this package against an independent method. Both are peptide purity batches measured twice, once by amino acid analysis and once by this package from the UV chromatogram.
130 injections quantified at both 214 and 280 nm. Because both numbers come from the same injection, everything about the sample cancels in their ratio and only the extinction model survives. The ratio turned out to depend on one residue and nothing else; see Recalibration of eps214.
aaa_hplc_compare/ holds the earlier comparison: 352 injections across four 2025 batches,
UV against amino acid analysis, in anonymised form. See
its README for the columns.
| Batch | n | median UV / AAA | R2 | Lin's concordance |
|---|---|---|---|---|
| 2025-01 | 52 | 1.16 | 0.17 | 0.34 |
| 2025-04 | 110 | 1.06 | 0.20 | 0.44 |
| 2025-05 | 124 | 1.02 | 0.57 | 0.74 |
| 2025-06 | 66 | 0.94 | 0.64 | 0.80 |
Two things to take from it. Each batch carries its own scale factor, spanning 0.94 to 1.16
here, so agreement should be judged within a batch rather than pooled across batches. And the
reference is not the fixed point it looks like: the reported amino acid analysis replicate
CV in this dataset has a median of 13.5 percent (interquartile 11.1 to 16.5), which is most
of the spread you see in scatter_by_batch.png. A UV number that sits 15 percent from an AAA
number is inside the reference method's own noise.
Not one of these four batches contained a tryptophan peptide, which is exactly why the problem in case 1 stayed invisible until 2026.
Each peak is integrated against a baseline drawn between its own two feet, not against zero after a global baseline has been subtracted from the whole trace.
The window is found by walking outward from the apex, and the walk runs to this peak's own foot. It is not stopped at the valley it shares with a neighbour: that valley sits partway up the flank, and anchoring the chord there lifted it a median 30 mAU off the baseline, cutting roughly a sixth of the area off the peak. Anchoring at the true foot instead leaves the chord a median 3.1 mAU above the local baseline, over 20 mAU on 2 runs of 47 rather than 32.
The shared valley is still honoured, as a bound on what gets integrated rather than on where the baseline is measured. Those are two different jobs and conflating them was the single largest error in this code. Every valley riding on the peak and every valley shared with a neighbour cuts the envelope, no cut is allowed to land on the apex itself, and the slice counted is the one holding this peak's own apex, so no two peaks claim the same stretch of trace and Area (%) stays a meaningful ratio. The level at each foot is fitted from the ten nearest points on the side away from the apex, so the fit never leans on the peak. Everything above the resulting chord is integrated, nothing below it.
The Integration detail panel in the app draws exactly that geometry, and
integrate_peak_against_endpoint_baseline() returns it if you want to draw it yourself.
Each foot is capped at eight peak widths from the apex, so the chord stays local. The global ALS correction still runs, to seed peak detection where a threshold needs a flat trace, but it no longer decides an area: the two baseline paths give areas within a median of 0.1 percent of each other.
sample_als_lambda = 4.5 and sample_als_p = 10^-6.5, chosen at the instrument by driving the
two sliders in the tuner against three test runs rather than by fitting a summary statistic.
Lambda is stiffness and p is how hard the fit is pushed under the data; the second is what keeps
a baseline off a peak, and tuning lambda alone cannot substitute for it.
Measured over 47 runs, against a lambda-only choice of 5.5 with p at 1e-4: the baseline's climb into the main peak falls from 2.51 mAU to 0.34, the worst it gets above the raw trace near the peak from 0.31 mAU to 0.02, and the offset it leaves in the first minute from 119 mAU to 78. Reported areas are unchanged, because they are taken against each peak's own chord.
The Baseline subtraction panel at the top of the app shows this fit, in two stacked panels sharing a time axis: the raw signal with the baseline drawn on it, and underneath, what is left after subtracting it. Its subtitle reports how far the baseline rises above the raw trace, which is the number that says whether it is cutting into anything.
The two panels are deliberately not superimposed. Until 0.7.7 this panel drew the baseline in the same axes as the corrected trace, and that comparison is meaningless: the corrected trace sits near zero while the baseline sits at whatever it is subtracting, so the baseline towers over it across the whole run and the panel reads as though the tool were destroying its own peaks. A baseline is above the signal it has already been subtracted from by exactly the amount it subtracted. The comparison that actually tells you something is against the raw trace it was fitted to, and that is the one drawn now. Measured that way over 47 production runs the baseline lies above the raw trace on 4.0 percent of points, by at most 0.31 mAU. The figure the panel prints is for the run in front of you, so it will differ from that; on the demo batch above it reads 3.9 percent and 2.55 mAU, the 2.55 coming from the negative injection artefact just after 2 minutes, which the fit cannot dive fast enough to follow.
The app used to take the blank-subtracting path whenever a blank folder was present, which is most batches. It does not any more, and the reason is that its baseline is not a baseline. On a production run the piecewise fit jumps 171 mAU between two adjacent points and dives to -171 at both ends of the chromatogram, where the plain ALS fit moves at most 0.8 mAU between points. It also over-subtracts the blank's own injector peak, leaving about -317 mAU at 3.3 min in every run.
None of that is worth paying for. Since each peak is integrated against its own chord, choosing
the blank path changes the reported area by a median of 1.6 percent. It remains available as
use_hybrid = TRUE on run_hplc_analysis_agilent().
At 280 nm the blank is never subtracted, even if you ask for it.
run_hplc_analysis_agilent() computes use_hybrid && signal_wavelength != 280, so an explicit
use_hybrid = TRUE is overridden at 280 nm. The reason is that there is nothing there to
subtract and the subtraction does harm. Measured over the eight blank runs of a production
batch, between 3 and 18 minutes, the blank baseline drifts about 2.8 mAU/min at 214 nm but
about 0.011 mAU/min at 280 nm, some 250 times less. Against hand-integrated areas on eight
well-resolved peaks during development, the hybrid path came out biased high on every one
(median +6.6 percent, worst +12.2 percent) while ALS alone stayed within 2.6 percent. The
override is unconditional rather than a changed default because the app sets use_hybrid from
"a blank folder exists in this batch", which is a fact about the plate layout and not a
scientific judgement.
The peptide sequence is resolved from three places, in order. The uploaded CSV map wins,
then SAMPLE.XML inside the .D folder, then the folder name. ChemStation truncates the
folder name at 40 characters, and it also caps the <Name> element in SAMPLE.XML at 40
characters. In the production batch tested, the longest folder basename is 42 characters
(40 plus .D) and the longest XML name is exactly 40; the XML resolved 57 of 69 runs against
47 for the folder name, was never shorter, and never disagreed. But five names sit at exactly
40 characters, and two of those sequences are visibly cut mid-peptide. For those, an uploaded
CSV map is the only way to get the real sequence in. The Download CSV map template button
writes a CSV prefilled with whatever the app resolved, so correcting them means editing two
cells rather than typing the whole plate.
Area (%) is against the ten peaks shown, not every peak detected. The denominator in
peak_area_percent() is the sum of the ranked peak areas that filter_top_peaks() returned,
which is the main peak plus its nine largest impurities. Dividing by every detected peak
instead sounds more rigorous but is not: on a 280 nm trace the detector reports dozens of
sub-mAU baseline features, and including them drags a peptide that is better than 95 percent
pure down to about 59 percent. The ranked set is the population a purity number is actually
about.
The on-screen peak table also drops anything before 6 minutes. That 6 minute floor is absolute and separate from the Min analyte RT slider, which is a percentage. A peak passing the slider can still be missing from the table if it elutes before 6 minutes.
In the app, the sidebar injection volume always wins. The value parsed from acq.txt is
used only when a script calls run_hplc_analysis_agilent(inj_vol_ml = NA). The app always
passes the sidebar value, so the file value is never consulted there.
| Control | What it does |
|---|---|
| Choose folder directory | Picks the batch folder. Agilent .D subfolders and Thermo *_UV_VIS_N.txt files are both detected. |
| Optional CSV map | Uploads sequence assignments. Highest priority sequence source. |
| Download CSV map template | Writes a CSV prefilled with the sequences the app resolved on its own. |
| Detection wavelength | 214 nm (Kuipers and Gruppen) or 280 nm (Edelhoch). Chooses the extracted trace on Agilent and the epsilon formula on both. |
| Thermo UV channel | The N in *_UV_VIS_N.txt. A file number, not a wavelength. |
| Dilution factor | Multiplies the reported concentration. Enter 10 for a 1:10 dilution. Non-finite or non-positive values fall back to 1. |
| Injection volume (uL) | Volume injected on column, default 100. Concentration scales inversely with it. Agilent only. |
| Load samples | Analyses every sample in the folder and fills the table. |
| Override blank | Forces one specific BLANK folder instead of the nearest preceding one. |
| Min analyte RT (% of run) | Drops peaks before this fraction of the run. Default 30. |
| Max analyte RT (% of run) | Drops peaks after this fraction of the run. Default 100, which keeps everything. |
| Reset integration | Clears the manual window for this sample and un-marks it as bad. |
| Mark as bad | Blanks the concentration for this sample and sets its status to Marked bad. |
| Previous / Next Sample | Navigation, synced with the table selection. |
| Download Results CSV | Exports the accumulated results. |
Both retention time bounds are percentages of run length rather than minutes, so one setting travels across methods of different duration. A single production batch already mixes 20.01 and 24.11 minute runs. If the maximum is set at or below the minimum, both revert to the defaults rather than reporting "no peaks detected" for the whole plate.
Manual integration. Brush horizontally across the lower peak plot to integrate a region by hand. The metrics box switches to that window's area and concentration and the results row is replaced. A hand-drawn window reports no Area (%), because it is not one of the ranked peaks and has no share of their summed area. Click Reset integration to return to the detector.
sample, rt, height, area, main_peak_area_pct, conc_uM_raw, sequence, status,
dilution_factor, conc_uM_final, wavelength_nm.
status distinguishes OK, NA (missing eps), Marked bad, No peaks detected and an
error string, so an empty concentration always carries its reason.
Column names are matched case-insensitively. Agilent runs are keyed on the MRMP id, Thermo runs on the plate well.
MRMP,Sequence
MRMP-00000001-001,SAMPLEPEPTIDEK
MRMP-00000001-002,TESTPEPTIDEAKWell,Peptide Sequence
P1-A1,SAMPLEPEPTIDEK
P1-B1,TESTPEPTIDEAKThe template the app writes carries MRMP, Well, Sequence and Run so that one file
serves both instruments. Extra columns are ignored.
| Setting | Recommended | Why |
|---|---|---|
| Detection wavelength | 214 nm | More than half of a typical peptide plate has no Trp or Tyr and cannot be quantified at 280 nm at all. |
| Min analyte RT | 30 percent | The shipped default. Keeps the injector and void region out of the peak table. |
| Max analyte RT | 80 percent | See below. |
| Injection volume | whatever the method injects | The app does not read it from the file. |
Why 80 percent for the maximum. Every method here ends with a column regeneration step at high organic, and the detector reports that step as a peak like any other. It appears in all runs including blanks, so it is instrument behaviour and not sample. Re-measured on this package at 214 nm over the 56 of 57 peptide runs of a production batch that complete, and across the ranked peaks the app actually reports, the latest genuine analyte peak sits at 74.6 percent of the run and the next ranked peak at 86.5 percent. Nothing at all falls in between, so any cutoff from about 75 to 86 percent separates them with no analyte lost, and 80 percent sits comfortably in the middle. Left at the default of 100 percent, the regeneration peak is the largest peak on many 280 nm runs and takes a share of Area (%) on all of them.
The default remains 100 percent so that upgrading the package never silently moves a number that has already been reported. Set it yourself.
- The Thermo path shares the wavelength control but uses it only for the epsilon formula. Selecting 280 nm while the Thermo channel selector is pointing at a 214 nm export will compute an Edelhoch eps280 against a 214 nm trace and report a concentration for it. Nothing in the code can detect this, because the Thermo reader never learns what wavelength it read.
- Two sequences in the production batch tested remain truncated at source, because
ChemStation caps the
SAMPLE.XMLname at the same 40 characters as the folder name. Only an uploaded CSV map can fix those. - The
acq.txtinjection volume parser requires a unit. The regex demandsuLormLafter the number, so an instrument that writes the number alone silently falls back to 100 uL. The instrument in the batch tested does write the unit and parses correctly, but the sidebar control exists so that this is never load bearing. - The reference wavelength is fixed at 360 nm in the app.
read_hplc_agilent()accepts asignal_refargument, but the app never sets it, so a.Dfolder without aSig=<wl>,Ref=360signal fails with "No matching DAD signal". Non-injection runs such as standby files fail this way by design. - Flow rate and pathlength are not exposed in the app, and default to 1 mL/min and 1 cm. A method running at a different flow rate needs the scripted interface.
- The hybrid baseline path can fail on individual runs. One of 58 runs in the batch tested
errors inside
baseline_hybrid_sm()withinvalid 'times' argument. The app catches it, and the row carries the error text rather than a number. - A peak sitting on a broad unresolved hump keeps the hump's level as its baseline. The chord lands a median 3.1 mAU above the local baseline across 47 runs, 6.2 at the 90th percentile, but on two runs it is about 20 mAU high because the peak is a shoulder on a rising feature and there is no flat stretch to anchor to. On one of those the cut also lands close enough to the apex that only part of the peak is integrated.
- Integration is not adjustable from the app. The foot walk, the valley prominence and the
ten point endpoint fit are constants in
R/integrate_peak_endpoint_baseline.R. A peak whose envelope cannot be walked silently keeps the area the detector's own trapezoid gave it. - The hybrid baseline path over-subtracts the injector peak, leaving a constant -317 mAU artefact at 3.28 min that the guard ramp does not cover, and returns concentrations 5 to 15 percent different from the ALS path on the same runs. See the blank-subtracting path leaves an injector artefact.
- The noise estimate is the standard deviation of the first 50 points of the trace. A run whose first 50 points are not baseline will have its detection threshold set wrongly.
- Test coverage is narrow.
tests/testthat/covers the extinction coefficient models and the calibration arithmetic, which are the numbers most likely to move. The readers, the baseline paths and the peak detection have no tests; they are verified by re-running production batches and comparing results. - The 214 nm coefficients are ours, not the paper's. Anyone reproducing a published
Kuipers and Gruppen number must call
published_epsilon_214(), notestimate_epsilon_214(). See Recalibration of eps214.
R/ package source
inst/shiny/app/app.R the Shiny front end, launched by run_hplc_app()
man/ roxygen-generated help pages
man/figures/ screenshots and the calibration figure used by this README
tests/testthat/ tests for the extinction models and the calibration
install.R dependency and package installer for a fresh clone
INSTALL.md step by step install guide for users new to R
NEWS.md version history
aaa_hplc_compare/ amino acid analysis versus UV-HPLC comparison, anonymised
peptide-calculator/ an Electron desktop peptide calculator, separate from the R package
aaa_hplc_compare/ and peptide-calculator/ are excluded from the built package by
.Rbuildignore.
The absorptivity models
-
Kuipers, B. J. H. and Gruppen, H. (2007). Prediction of molar extinction coefficients of proteins and peptides using UV absorption of the constituent amino acids at 214 nm to enable quantitative reverse phase high-performance liquid chromatography-mass spectrometry analysis. Journal of Agricultural and Food Chemistry 55(14), 5445-5451. doi:10.1021/jf070337l Source of the additive 214 nm model in
published_epsilon_214(): 923 per peptide bond, 2675 per non-N-terminal proline, the per-residue table, and the four special tripeptides. Measured in acetonitrile and formic acid, which is why it suits RP-HPLC eluent. -
Pace, C. N., Vajdos, F., Fee, L., Grimsley, G. and Gray, T. (1995). How to measure and predict the molar absorption coefficient of a protein. Protein Science 4(11), 2411-2423. doi:10.1002/pro.5560041120 Source of
estimate_epsilon_280(): 5500 per Trp and 1490 per Tyr. The 125 per disulfide is deliberately not implemented, because the disulfide state of a purity peptide is unknown and the term is about 2 percent of a single tryptophan. -
Edelhoch, H. (1967). Spectroscopic determination of tryptophan and tyrosine in proteins. Biochemistry 6(7), 1948-1954. doi:10.1021/bi00859a010 The original method that Pace and colleagues revised.
The signal processing
-
Eilers, P. H. C. and Boelens, H. F. M. (2005). Baseline correction with asymmetric least squares smoothing. Leiden University Medical Centre report. The ALS baseline behind
baseline_als()andbaseline_hybrid_sm(). -
Savitzky, A. and Golay, M. J. E. (1964). Smoothing and differentiation of data by simplified least squares procedures. Analytical Chemistry 36(8), 1627-1639. doi:10.1021/ac60214a047 The smoother
detect_peaks_on_smoothed()applies before peak picking. -
Bland, J. M. and Altman, D. G. (1986). Statistical methods for assessing agreement between two methods of clinical measurement. The Lancet 327(8476), 307-310. doi:10.1016/S0140-6736(86)90837-8 The difference plots in
aaa_hplc_compare/. -
Linnet, K. (1993). Evaluation of regression procedures for methods comparison studies. Clinical Chemistry 39(3), 424-432. Why the calibration in Recalibration of eps214 is fitted with errors allowed in both methods rather than by ordinary least squares.
The R packages this leans on
chromConverter reads the instrument files, baseline provides the ALS implementation,
pracma the peak finder and the trapezoidal integration, signal the Savitzky-Golay filter,
and shiny, shinyFiles and DT the front end. Run citation("chromConverter") and so on
for their preferred citations.
GPL-3. Copyright Peter Kubiniok.
If you use hplcAnalyzer, cite the software together with references 1 and 2 above:
Kubiniok, P. (2026). hplcAnalyzer: automated HPLC-UV analysis and peptide concentration estimation. R package version 0.7.8.
If you report a 214 nm concentration from it, say which coefficients you used. The shipped
estimate_epsilon_214() is not the published Kuipers and Gruppen value for tryptophan;
see Recalibration of eps214.
Contact: Peter Kubiniok, peterkubiniok@gmail.com