Dear MS-DIAL developers,
I am using MS-DIAL 4.9.22118 to analyze Waters MSE (DIA) data from a single traditional Chinese medicine formula sample. I encountered an issue when trying to export deconvoluted MS2 spectra for GNPS molecular networking.
Problem description:
- MS2Dec deconvolution works correctly in the GUI - I can see clean "Deconvoluted MS/MS spectrum" in the Peak spot window (e.g., only 2-3 major fragments like m/z 113.9574 and 149.0176 for precursor 158.956).
- However, when using File → Export → Peak list result → Spectra type: "deconvoluted", the exported .mgf file contains continuous ion series (e.g., m/z 60-159 for precursor 158.956), which appears to be centroid data rather than truly deconvoluted spectra.
- The "centroid" option exports m/z 50-1000, which is clearly raw full-scan data.
- The difference between "deconvoluted" and "centroid" exports is minimal - both contain many more fragments than what is shown in the GUI's Deconvoluted MS/MS spectrum view.
Data and settings:
- Data type: Waters Synapt G2-Si MSE (positive mode)
- Sample: Single sample (no alignment performed)
- MS-DIAL version: 4.9.22118
- Import mode: SWATH-MS + experiment file
- MS2Dec parameters: Sigma window 0.5, MS/MS abundance cut off 0, Exclude after precursor ion checked
What I've tried:
- Waters2mzML conversion → same issue
- Direct MS-DIAL import with SWATH-MS + experiment file → MS2Dec works in GUI but export fails
- Different export options (Molecular spectrum networking export, Peak list export) → all export raw/centroid data, not deconvoluted
- Checked .dcl and .pai2 files in project folder → binary format, cannot be read by external tools
Expected behavior:
The "deconvoluted" export option should produce .mgf files containing only the clean, deconvoluted fragments shown in the GUI's "Deconvoluted MS/MS spectrum" view (e.g., 2-10 major fragments per compound, not continuous ion series).
Questions:
- Is this a known bug in version 4.9.22118?
- Is single-sample MSE data fully supported for deconvoluted spectrum export?
- Are there any workarounds to export the truly deconvoluted spectra?
- What is the format of .dcl and .pai2 files? Is there any API or tool to extract deconvoluted spectra from them?
Thank you for your excellent work on MS-DIAL. I would greatly appreciate any guidance on this issue.
Best regards,
LZlulu
Dear MS-DIAL developers,
I am using MS-DIAL 4.9.22118 to analyze Waters MSE (DIA) data from a single traditional Chinese medicine formula sample. I encountered an issue when trying to export deconvoluted MS2 spectra for GNPS molecular networking.
Problem description:
Data and settings:
What I've tried:
Expected behavior:
The "deconvoluted" export option should produce .mgf files containing only the clean, deconvoluted fragments shown in the GUI's "Deconvoluted MS/MS spectrum" view (e.g., 2-10 major fragments per compound, not continuous ion series).
Questions:
Thank you for your excellent work on MS-DIAL. I would greatly appreciate any guidance on this issue.
Best regards,
LZlulu